• 1 (1)

Medical Isolation Power Supply System

53

Application

Medical IT system insulation monitoring device and fault location system is suitable for hospital operating room, ICU (CCU) and other important places, and can provide a safe place for such continuous, reliable power solutions.

Structure

1

Main Functions

1)Primary connection diagram and field distribution display

1
2

2)Real-time data acquisition and display

3

3)Fault Alarm

4

4)Remote parameter setting and inquiry

5

5)Graphic display

Display the insulation status, load status, and the isolation transformer temperature of each set of isolated power systems in graph form.

Product Selection

1

AITR Series Medical isolation transformer

Isolating transformers for medical IT systems

Double insulation treatment, Electrostatic shielding layer

PT100 temperature sensor

Rated power: 3.15…10KVA

CE mark

2

AID10 Alarm and Display Instrument

Insulation monitoring for medical IT systems

Load and temperature monitoring of isolation transformer

Device disconnection monitoring

LED indicators: On, Insulation, Overload, Overtemp

Alarm relay output

Test button

RS485,MODBUS-RTU

3

AIM-M100 Medical Intelligent Insulation Monitor

Insulation monitoring for medical IT systems

Load and temperature monitoring of isolation transformer

Device disconnection monitoring

LED indicators: On, Insulation, Overload, Overtemp

Alarm relay output

Test button

RS485,MODBUS-RTU

4

AIM-M200 Medical Intelligent Insulation Monitor

Insulation monitoring for medical IT systems

Load and temperature monitoring of isolation transformer

Device disconnection monitoring

LED indicators: On, Insulation, Overload, Overtemp

Alarm relay output

Test button

Insulation fault location

RS485 (MODBUS-RTU) + CAN (custom protocol)

5

AID10 Alarm and Display Instrument

Remote monitoring

LED indicators:On, Insulation, Overload, Overtemp

Mute button

Test button

Power supply: DC24V

RS485,MODBUS-RTU

6

AID120 Alarm and Display Instrument

Remote monitoring

LED indicators:On, Insulation, Overload, Overtemp

Mute button

Test button

Power supply: DC24V

RS485,MODBUS-RTU

7

AID150 Alarm and Display Instrument

Remote monitoring

LED indicators:On, Insulation, Overload, Overtemp

Mute button

Test button

Power supply: DC24V

RS485,MODBUS-RTU

8

ACLP10-24 DC Power Supply

Power Supply to AID series

AC 220V input, DC 24V output

Completely Isolated Linear Transformer

Power-on indicator

9

ASG150 Signal Generator

Generating Fault Location Signal

Detection of L1 and L2 disconnection

LED indicators: On, Comm, L1, L2, Fault

CAN bus, custom protocol

Power supply: DC24V

10

AIL150-4/-8 Insulation Fault Locator

Insulation fault location in AC IT systems

AIL150-4: 4 measuring channels

AIL150-8: 4 measuring channels

LED indicators: On, Comm, L1…L8

CAN bus, custom protocol

Power supply: DC24V

Typical Solution

1)Four-piece set

1

2)Five-piece set

2

3)Seven-piece set

3

Photos on site

1