The AIM-M200 monitors the insulation resistance of medical IT system, At the same time, the load current and temperature of the IT system transformer is monitored. In combination with ASG100, AIL100 series insulation fault locators and the appropriate measuring current transformers, the AIM-M200 is designed to support insulation fault location.
Technical Parameters of AIM-M200 Medical Intelligent Insulation Monitoring
AUX Power |
Voltage |
DC 18~36V |
Temperature monitoring |
Thermal resistor |
Pt100 |
Power consumption |
≤3W |
Measuring range |
-50~+200℃ |
||
Insulation monitoring |
Resistance measuring range |
15~999kΩ |
Alarm value range |
0~+200℃ |
|
Response value |
50~999kΩ |
Alarm output |
Output mode |
Relays |
|
Relative uncertainty |
±10%,±10K |
Contact rating |
AC 250V/3A DC 30V/3A |
||
Response time |
≤3s |
Environment |
Operating temperature |
-10~+55℃ |
|
Permissible system leakage capacitance Ce |
≤5uF |
Transport temperature |
-25~+70℃ |
||
Measuring voltage Um |
≤12V |
Storage temperature |
-25~+70℃ |
||
Measuring current Im |
≤50uA |
Relative humidity |
5~95%,No condensation |
||
Impedance Zi |
≥200kΩ |
|
Altitude |
≤2500m |
|
Internal DC resistance Ri |
≥240kΩ |
IP degree |
IP30 |
||
Permissible extraneous DC voltage Ufg |
≤DC280V |
Rated impulse voltage / pollution degree |
4KV/Ⅲ |
||
Load current monitoring |
Measuring Value |
2.1~50A |
EMC/EMR |
IEC 61326-2-4 |
|
Alarm Value |
5~50A |
Communication protocol |
CAN, customize |
||
Measuring accuracy |
±5% |
RS485, Modbus-RTU |