The insulation fault locator AIL150-4/-8 in conjunction with AIM-M200 medical intelligent insulation monitor and ASG150 test signal generator, are designed for insulation fault location in medical IT systems. The locating current generated by ASG150 are detected using the integrated measuring current transformers and evaluated by the insulation fault locators.
Technical Parameters
Item |
Technical parameters |
||
ASG150 |
AIL150-4 / 8 |
||
Auxiliary power supply |
Voltage |
DC 24V |
DC 18~36V |
Maximum power consumption |
≤2VA |
||
Monitored system |
Rated voltage |
0~242 VAC |
|
Rated frequency |
45~60Hz |
||
Locating current |
<1mA r.m.s. |
— |
|
Fault locating |
Maximum circuit |
— |
4、8 |
Response time |
— |
≤5s |
|
Communication |
Mode |
CAN |
|
Protocol |
customize |